PC950 pH/Conductivity Meter, Benchtop

Instrument Features

  • Advanced, Intuitive Interface — Plain-language menus and graphic screen prompts eliminate the need for extensive manual reading or training. Multi-language support includes English, German, Spanish, and Chinese.

  • Dual-Channel Simultaneous Measurement — Measures pH/mV(ORP) and conductivity/TDS/salinity at the same time for maximum efficiency.

  • GLP Data Management — Manual/automatic data logger with 200 sets of GLP-compliant data storage containing calibration data and time stamps. Export data via USB to PC.

  • Apera TestBench Included — Integrated workstation with a flexible electrode holder, intelligent magnetic stirrer, and buffer organizers — dramatically streamlines calibration and measurement workflows.

pH/mV (ORP) Measurement
  • LabSen® 211 pH Electrode — Features a low-impedance glass membrane and long-life reference system. Supports 0–100°C ATC via the MP500 temperature probe. Compatible with TRIS buffers.
  • Flexible Calibration Options — 1 to 3 point auto. calibration recognizing standard buffers (1.68 / 4.00 / 7.00 / 10.01 / 12.45) or user-defined standards. Automatic electrode slope display and calibration self-diagnosis.

  • Broad Compatibility — Standard BNC pH connector & RCA temperature connector (NTC 30KΩ). Compatible with all Swiss LabSen electrodes for specialized samples.

Conductivity/TDS/Salinity/Resistivity Measurement
  • 2401T-F Conductivity Electrode — glass-body platinum black sensor delivers high-accuracy measurements from 0 to 200 mS/cm at ±1% F.S accuracy.

  • Flexible Conductivity Calibration — 1 to 4 point auto. calibration with standard solutions (84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm) or user-defined standards.

  • Quick Parameter Switching — One-tap conversion among conductivity, TDS, and salinity.

  • Broad Compatibility — Standard BNC conductivity connector; compatible with 2310T-F (up to 2000 mS/cm) and DJS-0.1-F for high-purity water. Uses RCA temperature connector (NTC 30KΩ).

Browse our catalogue

SKU: PC950 Category: Tags: , , Brand:
Parameter Settings Menu
Parameter Menu Options Factory Default
1.0 pH
1.1 pH Buffer Selection CH / USA / NIST / User USA
1.2 Resolution 0.01 / 0.1 0.01
1.3 Last Calibration Data View —
1.4 Restore Factory Default Yes / No No
2.0 Conductivity / TDS
2.1 Cell Constant 10 / 1.0 / 0.1 1.0
2.2 CAL Solution Selection Standard / CN / User Standard
2.3 Last Calibration Data View —
2.4 Reference Temperature 15–30 °C 25 °C
2.5 Temp Compensation Factor 0.00–10.00% 2.0%
2.6 TDS Factor 0.40–1.00 0.71
2.7 Restore Factory Default Yes / No No
3.0 Configuration
3.1 Reading with HOLD On / Off —
3.2 Auto Timing Data Log Manual / Automatically Manual
3.3 Temperature Unit °C / °F —
3.4 Delete Saved Data Yes / No No
3.5 Language Selection English / Deutsch / Español / Chinese English
3.6 Date Format Y-M-D —
3.7 Time Format 24-hour / 12-hour —
3.8 Date Setting YYYY-MM-DD —
3.9 Time Setting HH:MM:SS —

 

pH Measuring range 0 to 14.00
Resolution 0,1/0,01
Accuracy ±0.01 ± 1 digit
Calibration Automatic 1 to 3-point calibration
Temperature compensation 0 to 100 °C (automatic or manual)
mV Measuring range ±2000 mV
Resolution 1 mV
Accuracy 0.1 % F.S. ± 1 digit
Conductivity Measuring ranges Total: 0 to 200 mS/cm 

0 to 19.99 μS/cm

20.0 to 199.9 μS/cm

200 to 1999 μS/cm

2.00 to 19.99 mS/cm

20.0 to 199.9 mS/cm

 

TDS: 0 to 100 g/l

Salinity: 0 to 100 ppt

Resolution 0.01 μS/cm; 0.1 μS/cm; 1 μS/cm
Accuracy ±1.0 % F.S. ± 1 digit
Cell constant 0.1/1/10 cm-1
Temperature compensation 0 to 50 °C (automatic or manual)
temperature Measuring range 0 to 100 °C
Resolution 0,1 °C
Accuracy ±0.5 °C ± 1 digit
Other Electrode connections pH/conductivity: BNC; 

Temperature: Cinch

Data storage 200 groups
Data content Numbering, measured value, temperature
Data output USB
Power supply unit DC 9 V/600 mA
Housing protection class IP54 (protection against dust and splash water)
Dimensions 360 × 235 × 400 mm
Weight 1.7 kg